Cleaning and maintenance of your InnoSpire Go

After each use:

    • Pour away residual medication from the medication chamber
    • Press the mouthpiece assembly release button to separate the mouthpiece assembly from the handset
    • Rinse the mouthpiece assembly thoroughly under running tap water
    • Shake off excess water and allow to air dry fully before storing

Daily cleaning:

    • Wash the mouthpiece assembly by hand in a bowl of warm soapy water (liquid dishwashing soap) for 2 minutes
    • Rinse the mouthpiece assembly thoroughly under running tap water
    • Shake off excess water and allow to air dry fully before storing

Weekly disinfection

    • CAUTION: Risk of scalding. Use care around boiling water and in handling hot parts
    • Prior to disinfection, ensure all parts are visibly clean and free from dirt/debris
    • Boil the mouthpiece assembly in water for 10 minutes. Ensure the medication chamber lid is open and there is enough water in the pan to prevent the mouthpiece assembly from touching the bottom or boiling dry. Shake off excess water and allow to air dry fully before storing
      OR
    • Immerse the mouthpiece assembly in a disinfectant of the Gluteraldehyde group (testing performed with Korsolex Extra, 4% for 15 minutes). Rinse thoroughly according to manufacturers instructions

Mask cleaning:

    • Wipe your handset clean with a clean damp cloth weekly or as needed. Do not use other cleaning methods or solutions

Handset cleaning:

    • If using a mask and mask adapter, once a week: agitate for 2 minutes or soak for 10 minutes, in warm soapy water (liquid dishwashing soap), rinse thoroughly under running tap water and air dry fully
    • After each cleaning procedure visually inspect for any remaining dirt/debris, if any remains repeat the procedure. Visually inspect for any moisture before storing

Maintenance:

    • Mouthpiece assembly can be cleaned (boiled/disinfected) up to 52 times over 12 months. To keep your InnoSpire Go working at the optimum level you should replace the mouthpiece assembly and mask, if used, every 12 months as this is a consumable part